Care and Use of the Bright Field Microscope

Malaria parasite

Mycobacterium tuberculosis

Page 3: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Importance of the Microscope

• Important for hematology, microbiology, TB, and malaria testing

• Compound microscope used in bacteriology, biology, and medicine to examine minute objects such as bacteria, other unicellular organisms, and plant and animal cells and tissue

• Advances in fluorochrome stains and monoclonal antibody techniques caused growth in use of fluorescence microscopy in both biomedical analysis and cell biology

Page 4: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Type of microscope

• Classification though light application

– Electron (vacuum)

• SEM and TEM

– laser

• Confocal

– Light

• Compound and stereo

Page 5: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Basic Components of the Microscope and their functions...

Power switch

Light intensity control

Condenser

Stage

Objectives

Eyepiece lens

Stage motion

control knobs

Course and fine

adjustments

knobs

Field diaphragm ring

Aperture diaphragm

Page 6: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Microscope Work Principle

• Manipulation of light

• Physical property of light

• Lens

• Air or vacuum

Page 7: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Example of Microscope

Page 8: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Example (continued)

Page 9: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Care and Maintenance of the Microscope

• Good preventive maintenance and care includes:

• Regular cleaning of oculars and objectives

• Avoid damaging oculars and other optics with eye make-up or other debris

• Careful handling to avoid abrupt motions

• Protect from direct sunlight, high temperature, humidity, dust and vibration

• Use appropriate materials to clean the lenses

• Cover when not in use with vinyl or plastic dust cover

Page 10: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Cleaning the Microscope

Routine Cleaning Supplies:

• Commercial lens tissue for optics

• Caution: Do not use paper towels or other rough paper products

• Cotton swabs with wooden shaft (optics)

• 70% isopropyl alcohol

• Dilute methanol is satisfactory

• Mild detergent and soft cloth for stage and base of microscope

Page 11: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Cleaning Oculars and Objectives

• Unplug the microscope

• Wash hands

• Remove dust from optical glass surfaces

• Carefully remove eyepieces, objectives, condenser, and filters–one at a time

• Excessive rubbing can cause damage to iridescent coating on lens

• Clean and replace as completed

• Do Not take eyepiece or objectives apart

Page 12: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

• Unplug microscope and allow bulb to cool

• Carefully place microscope on its side

• Open bulb house; use tissue to remove bulb

• Use tissue (to avoid fingerprints) to pick up new bulb

• Insert new bulb and close bulb house

Replacing Microscope Bulb

Page 13: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

• Plug in microscope and turn on illuminator. Rotate nosepiece to lock 10X objective in place

• Place smear on stage and center it under the 10X objective

• Open the field diaphragm all the way and close condenser diaphragm all the way

• Move up (rack up) stage to its highest position

• Adjust the oculars for interpupillary distance so that only one circle of light is seen

• Rack up condenser as high as possible

Setting the Koehler Illumination

Page 14: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

• Close field diaphragm half way and focus smear at 10X

• Close field diaphragm until diameter of illuminated image is smaller than the field of view

• Lower condenser with positioning knob until you have a sharp, focused image of the edges of the field diaphragm

• Adjust condenser using centering screws so that the circle of light is centered in field

• Open field diaphragm until illuminated image is just larger than the field of view. If more light is needed, use the transformer.

• Koehler illumination is now set. It is important not to move the condenser up or down or change the field diaphragm.

Setting the Koehler Illumination (continued)

Page 15: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Operation of the Microscope – Examining Smears

• Put smear on stage and center it under the 10X objective

• Adjust intensity of the light to a comfortable level with the transformer

• Open condenser diaphragm about 70% to achieve a good balance of resolution and contrast

• Adjust oculars for inter pupillary distance so that when looking with both eyes only one circle of light is seen

Page 16: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Examining Smears (continued)

• Adjust sharpness of image by moving adjustment ring on adjustable ocular

• Once 10X focus is achieved, rotate nosepiece so that the 40X objective is in place

• Readjust the intensity of light to a comfortable level using the transformer

• Use the fine adjustment knob to focus up and down through the different planes of the field

Page 17: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Microscope Problems – Troubleshooting 1

Problem: Black Field

Possible Causes:

• Microscope not plugged in

• Power not available at outlet

• Illuminator not turned on

• Bulb burned out

• Objective not clicked into place

• Condenser too low with diaphragms closed

Page 18: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Microscope Problems – Troubleshooting 2

Problem: Field only partially illuminated

Possible Causes:

• Objective not clicked into position

• Condenser not centered correctly

• Condenser too low

• Field diaphragms closed too much

Page 19: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Problem: Difficulty focusing with 10X

objective

Possible Causes:

• Wrong objective in place

• Objective not screwed into place

• Not in correct plane of focus

Microscope Problems – Troubleshooting 3

Page 20: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Microscope Problems – Troubleshooting 4

Problem: Difficulty focusing with 40X objective Possible Causes: • Not in correct plane of focus • Not initially focused at 10X

Page 21: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Microscope Problems – Troubleshooting 5

Problem: Blurry image at 10X or 40X

Possible Causes:

• Dirty objective

• Dirty slide

• Dirty coverslip

Problem: Ground glass appearance

Possible Causes:

• Condenser too high

Page 22: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Storing The Microscope

• Let the microscope cool down before

storing

• Place the microscope at cabinet that is dry,

low lighted, and covered.

• Cabinet best condition is when it is free

from micro and major vibration

Page 23: KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN …KURSUS PEMBANTU MAKMAL KEMENTERIAN PENDIDIKAN MALAYSIA MOHD HANAFY BIN GAUSMIAN PEGAWAI SAINS UNIVERSITI KEBANGSAAN MALAYSIA . Care

Transferring Microscope

• Should a microscope to be transfer from one

location to the other.

• Carry the microscope with caution by

holding to its arm while the other hand

place at the bottom of the microscope.