[ieee 2012 international conference on biomedical engineering (icobe) - penang, malaysia...
TRANSCRIPT
Simulation of Electromagnetic Field (EM) Focusing
Capability on Biological Tissue through the
Application of C-type Excitation Coil Screen
Zulkarnay Zakaria, Noor Alia Mohd Zain, Azian
Azamimi Abdullah, Sofi Yahya
Biomedical Electronic Engineering Dept
School of Mechatronic Engineering
Universiti Malaysia Perlis
Arau, Perlis, Malaysia
Ruzairi Abdul Rahim, Muhammad Saiful Badri
Mansor, Abdul Rahim Mohd Disar Dept. of Control and Instrumentation Engineering
Faculty of Electrical Engineering
Universiti Teknologi Malaysia
Skudai, Johor, Malaysia
Abstract – In Magnetic Induction Tomography (MIT) or
Electromagnetic Therapy system, excitation coil is applied with
AC source in generating the electromagnetic (EM) field which
then propagate and penetrate the object located in the region of
interest (ROI). However, instead to the target object, the fields
also propagate around the coil and create interference to the
nearby circuit which contributes noise to the system while at the
same time wasting the energy. This paper is focusing on the use
of C-type excitation coil screen in focusing the EM field to the
ROI and measure the penetration depth when different
frequency is applied to the generation system.
Keywords-component; Magnetic induction; excitation coil;
tomography; Electromagnetic therapy; screen.
I. INTRODUCTION
Magnetic Induction Tomography (MIT) [1] and
Electromagnetic Therapy [2] system are the example of
systems which used electromagnetic (EM) field as a source in
their applications. The EM fields generate by excitation coil,
then will propagate and penetrate the object located at the
region of interest (ROI). Due to general excitation coil
(without screen), while EM propagate to the ROI, there are
also some portion of the field which surround the coil and
interfere the system. Instead of contribute noise to the system
through the generation of heat, this phenomenon also can be
considered as wasting of EM energy.
Based on that motivation, Zakaria et al. (2011) [3],
Stawicki et al.(2009) [4] and Barba et al. (2009) [5] had done
their study on the excitation coil screen design in focusing the
EM field to the ROI while minimizing the interference effects
of the field to the electronic circuit system. Stawicki and
Barba had reported the used of cone-type screen while Zakaria
had compared the performance of C-type and Cone-type
screen in his research. Zakaria also had reported that both C-
type and Cone-type excitation coil screen are capable in
focusing the EM field to the ROI and at the same time
reducing the scattered EM field to the nearby system
especially the electronic circuitry.
II. FORMULATION OF EM FIELD
All EM field are assumed fulfilled the state of the art of
Maxwell’s equation that are:
∇ × � = −jω� (1)
∇ × = (� + jωε)� (2)
∇ ∙ � = � (3)
∇ ∙ � = 0 (4)
When EM field penetrate an object, where in this case is
biological tissue, eddy current is induced in the biological
tissue itself due to passive electrical properties (conductivity,
σ; permittivity, ε; and permeability,µ) of the tissue. However
in most cases of biological tissue cases, conductivity is
dominant compare to others, hence the equation becomes:
� = �� = −��(��� +���) (5)
where φ is scalar potential while A is vector potential and J is
the current density in a material.
Magnetic field strength at any point in the ROI due to
induced eddy current can be calculated through Biot-Savart
law that is:
� = ��∑ ���× !" #$ (6)
Base on the equation, the magnitude of the magnetic field
strength, � is also depends on the area or volume covered by
the field since dV is the small element of a volume. Deeper
penetration normally will involve volume. Related to that, the
penetration of the field inside the material (biological tissue) is
depends on the frequency used and is given by the formula:
% = &�' (
)*+*,-./0 ( 7)
2012 International Conference on Biomedical Engineering (ICoBE),27-28 February 2012,Penang
978-1-4577-1991-2/12/$26.00 ©2011 IEEE 598
Where d is penetration depth, ω is frequency, µ is
permeability, ε0 is free space permittivity, εr is relative
permittivity and tan δ is loss factor.
Related to that, thermal effect is another phenomenon exists
in biological tissue in relation with EM field. Wang et al.
(2009) [6] explained that, thermal effects are indirect
interaction with biological tissues due to the high RF
radiation, where in this condition, electromagnetic field
generates the heat and causing biological effect. The
absorption of EM energy within the biological tissues causing
the biological effect due to temperature rising. Tissue exposed
to EM fields will continue rising with temperature until the
heat absorption rate is balanced with the rate at which it is
dissipated. The temperature dissipation is due to conduction
with other tissue types, convection through blood perfusion
and radiation to the surroundings. This heating concept has
been applied as therapeutic technique in curing cancer [7]. On
the other hand, under localized (partial body) near-field
exposure conditions, the internal fields decay exponentially
with distance from the exposed external surface, and the rate
of decay depends on the conductivity of the tissue [8].
III. EXPERIMENTAL SETUPS
For this research, C-type excitation coil designs had been
considered with frequency are set to 0.5 kHz, 1 kHz, 5 kHz,
20 kHz, 50 kHz and 100 kHz which are the acceptable range
of a biological tissue imaging [9], number of turn is 10 with 1
A current flow and 10 AWG wire. The properties of skin and
muscle represent biological tissue are as in Table 1 [10]. The
thickness of skin and muscle are approximately 0.2 cm and 4.0
cm respectively.
Figure 1. C-type excitation coil design with biological tissue model in FEMM
environment
TABLE 1. ELECTRICAL PROPERTIES OF SKIN AND MUSCLE
BASED ON FREQUENCY
Frequency
(kHz)
Skin (Dry) Muscle
Conductivitty
(S/m)
Permittivitty Conductivitty
(S/m)
Permittivitty
100 0.00045128 1119.2 0.36 8089
50 0.00027309 1126.8 0.35 10094
20 0.00021417 1131.7 0.34 15521
5 0.00020117 1134.6 0.33669 52349
1 0.00020006 1135.6 0.32115 434930
0.5 0.00020002 1135.8 0.30972 1087500
IV. RESULTS AND DISCUSSIONS
Based on the simulation results, it is found that even with the
used of the electromagnetic screen on the excitaion coil, the
scatered field is still happened at the above, bottom and back
region of the screen when the appied frequency is 5 kHz and
below as shown in Fig. 3-5. However scattering effect is totally
eliminated when frequency 20 kHz and above are applied. In
term of magnetic field exposure to the tissue, 0.5 kHz
frequency has proven of providing more higher magnitude of
magnetic field compare to all others as in Fig. 9.
Figure 2. Mesh of the experimental model
Figure 3. Magnetic field strength distribution at 500 Hz
Coil
Aluminum
screen
A B C
0 1.2 cm 8.8 cm
599
Figure 4. Magnetic field strength distribution at 1 kHz
Figure 5. Magnetic field strength distribution at 5 kHz
Figure 6. Magnetic field strength distribution at 20 kHz
Figure 7. Magnetic field strength distribution at 50 kHz
Figure 8. Magnetic field strength distribution at 100 kHz
Figure 9. Magnetic field strength of distribution from 500 Hz to 100 kHz
On the penetration capability, higher frequency will have more
penetration depth with wide coverage of biological tissue area.
For a MIT system with the aim of imaging the whole body, the
frequency of higher than 100 kHz should be applied for the
purpose of giving more penetration depth in the body with at
least half of the body diameter. However for the
electromagnetic therapy purposes, application of frequency
within the range of 5 kHz - 20 kHz is enough as long as the
scattering effect is eliminated.
CONCLUSION
C-type excitation coil screen has been proven of minimizing
the scattered EM field and at certain frequency apply, it may
fully eliminating the scattering effects at the back, above and
below the screen area. For next stages, the design of the screen
needs to undergo some modifications, thus providing the
0.00E+00
5.00E-05
1.00E-04
1.50E-04
2.00E-04
2.50E-04
3.00E-04
3.50E-04
0.0
0
0.6
0
1.2
1
1.8
1
2.4
2
3.0
2
3.6
2
4.2
3
4.8
3
5.4
4
6.0
4
6.6
4
7.2
5
7.8
5
8.4
6
Ma
gn
eti
c fi
eld
str
en
gth
(B
), T
esl
a
B (Tesla) vs Distance (cm)
0.5k
1k
5k
20k
50k
100k
600
scattering EM field eliminating even in the application of very
low frequency.
.
ACKNOWLEDGMENT
This work is supported by the FRGS grant 9003-00248 by the
Ministry of Higher Education of Malaysia and Science fund
grant 06-01-06-SF0889.
REFERENCES
[1] A.J. Peyton, Z.Z. Yu, G. Lyon, S. Al-Zeibak, J. Ferreira, J. Velez, F.
Linhares, A.R. Borges, H.L. Xiong, N.H. Saunders, and M.S. Beck, “An
overview of electromagnetic inductance tomography : description of three different systems,” Measurement Science and Technology, vol. 7,
1996, p. 261-271
[2] Zulkarnay Zakaria, Mohd Fahajumi Jumaah, Mohd Saiful Badri Mansor,
Khairi Mat Daud, Mohd Hafiz Fazalul Rahiman, Ruzairi Abdul Rahim
and Syed Mustafa Kamal Syed Aman (2011), “Initial Results On
Medium Frequency Electromagnetic Field Penetration In Biological Soft
Tissue”, Jurnal Teknologi, Vol. 54 Januari 2011: 69-77
[3] Zulkarnay Zakaria, Muhamad Hafiz Bin Hussin, Ruzairi Abdul Rahim,
Nur Farahiyah Mohammad, Azian Azamimi Abdullah, Sazali Yaacob and Syed Mustafa Kamal Syed Aman, “Performance Comparisons of
New Excitation Coil Design Aspects in Magnetic Induction
Tomography (MIT) Applications”, 2nd International Conference on
Intelligent Systems, Modelling and Simulation, ISMS2011, 25 – 27
January 2011, Phnom Penh, Cambodia.
[4] K. Stawicki, S. Gratkowski, M. Komorowski, and T. Pietrusewicz, “A New Transducer for Magnetic Induction Tomography,” IEEE
Transactions on Magnetics, vol. 45, no. 3, pp. 1832-1835, Mar. 2009.
[5] P. D. Barba, M. E. Mognaschi, R. Palka, and A. Savini, “Optimization
of the MIT Field Exciter by a Multiobjective Design,” IEEE
Transactions on Magnetics, vol. 45, no. 3, pp. 1530-1533, Mar. 2009.
[6] Zhao Wang, “Electromagnetic Field Interaction with Biological Tissues
and Cells”, Ph.D. Dissertation, Dept. School of Electronic Engineering
and Computer Science, Queen Mary, University of London, April 2009.
[7] J. Mackerle, “Finite element analyses and simulations in biomedicine: a bibliography (1985-1999),” Engineering Computations, vol. 17, no. 7,
pp. 813-856, 2000.
[8] The Institute of Electrical and Electronics Engineers, Inc, “IEEE
Recommended Practice for Measurements and Computations of Radio
Frequency Electromagnetic Fields With Respect to HumanExposure to
Such Fields, 100 kHz–300 GHz”, United State: .,IEEE-SA Standards Board., 13 January 2003
[9] Hermann Scharfetter, Helmut K Lackner and Javier Rosell (2001): ,
“Magnetic induction tomography: hardware for multi-frequency
measurements in biological tissues”, Physiological Measurements,
Volume 22, Number 1, pp. 131–146.
[10] Dielectric properties of body tissue, http://niremf.iroe.fi.cnr.it/tissprop; retrieved 15 Jan 2012
601